ToggleKITOZER MRI Shielded Room Installation & Commissioning Guide
Brand: KITOZER Series: KTR-MR Category: EMI/RF Shielding > Special Shielded Chambers > MRI Shielded Chamber
Turnkey delivery for radiology departments, research institutes and industrial NMR sites — from survey to shielding-effectiveness certification.
1. Overview
Magnetic Resonance Imaging (MRI) relies on a strong static field, gradient field and RF pulses to produce high-resolution soft-tissue images. It is highly sensitive to external interference: broadcast, cellular, radar and power-line signals, together with mechanical vibration, produce streaks and artifacts; at the same time, the scanner’s own RF energy and fringe field can disrupt nearby electronics and pose a hazard to patients with pacemakers or implanted defibrillators.
The KITOZER KTR-MR MRI shielded room is a six-sided welded Faraday cage that delivers bidirectional shielding — blocking external RF, power-frequency magnetic fields and vibration from reaching the coil, while containing the scanner’s own emissions and keeping the 5-gauss (0.5 mT) line inside the magnet room. Standard builds use 0.3-0.5 mm T2 copper sheet; cost-optimized builds use 1.0-1.5 mm galvanized steel; high-field scanners can add a permalloy passive magnetic shielding layer. The series covers 0.2 T to 7.0 T clinical and research systems.
2. Key Features
RF shielding (Faraday cage)
- Fully welded six-sided copper shell; every seam is 100% continuity-tested. Skin depth of copper at 10 MHz is ~0.02 mm, so 0.2 mm copper already yields >90 dB attenuation.
- SE >=100 dB (Grade A) / >=90 dB (B) / >=80 dB (C), measured per GB/T 12190 (equivalent to IEEE 299 / ASTM E1851).
Static & gradient magnetic shielding
- For 3.0 T and above, layered pure-iron plus permalloy passive shielding constrains the 5-gauss line inside the magnet room.
- The copper shell also provides passive gradient shielding, suppressing eddy currents induced in the building structure.
Specialized penetrations
- Shielded door: double-row beryllium-copper fingerstock, manual swing or motorized sliding, opening force <=150 N.
- Shielded viewport: double toughened glass with embedded copper mesh, light transmission >=80%.
- Honeycomb vent panels: 300×300 / 500×500 mm waveguide-below-cutoff vents.
- Power & signal filters: insertion loss >=60 dB, current range 10 A-630 A.
3. Models & Specifications
| Model | Scanner | Construction | Grade | Typical use |
|---|---|---|---|---|
| KTR-MR-1.5T-T-B | 1.5 T | Welded copper | >=90 dB | Standard hospital MRI |
| KTR-MR-3.0T-T-A | 3.0 T | Welded copper | >=100 dB | Tertiary imaging center |
| KTR-MR-3.0T-C-A | 3.0 T | Copper + permalloy | >=100 dB + mag-shield | Near public corridors / precision instruments |
| KTR-MR-7.0T-T-A | 7.0 T research | Heavy copper + multi-layer mag-shield | >=100 dB + strong mag-shield | Research / animal MRI |
| KTR-MR-G-1.5T | 1.5 T | Welded galvanized steel | >=80 dB | Budget-sensitive, moderate LF requirement |
Key parameters
| Item | Value |
|---|---|
| RF shielding effectiveness | >=100 dB (10 kHz-3 GHz, Grade A) |
| Magnetic shielding | 5-gauss line inside magnet room (optional) |
| Shell material | 0.3-0.5 mm T2 copper / 1.0-1.5 mm galvanized steel |
| Floor live load | >=20 kN/m² (upgradeable) |
| Seismic | Intensity 8 (GB 50011) |
| Earth resistance | Dedicated earth <=1 Ω; high-demand <=0.5 Ω |
| Room noise | <=65 dB(A) during gradient operation |
| Operating environment | 5-40 °C, <=85% RH non-condensing |
| Design life | >=30 years |
4. Reference Price (typical 1.5 T room)
KTR-MR-1.5T-T-B standard copper room (~6.5×5.0x3.2 m, incl. door/viewport/vent/filters/earth): RMB 850,000-1,100,000 / room
KTR-MR-3.0T-T-A high-grade room (equipment door, quench-pipe waveguide, power/signal filters): RMB 1,300,000-1,800,000 / room
KTR-MR-3.0T-C-A composite magnetic-shielded room (permalloy layer): RMB 2,200,000-3,000,000 / room
Reference prices only; final quotation applies. Based on exchange rate as of October 2026 (1 USD = 7.2 CNY), equivalents are roughly USD 118k-153k / 180k-250k / 305k-415k per room. Price covers material manufacturing, site installation, SE tuning and third-party report support; excludes the MRI scanner itself, medical gas, finish flooring and the outdoor quench-pipe segment.
5. Applications
- Hospital radiology / imaging departments (0.2 T-3.0 T clinical).
- Research institutes (3.0 T / 7.0 T human or preclinical MRI labs).
- Industrial NMR / non-destructive testing enclosures.
- Electron microscopy and precision-instrument shielded rooms.
6. Installation & Commissioning
Pre-site
- Survey clear dimensions, floor load (>=15 kN/m² for 3.0 T areas), flatness <=3 mm/2 m, moisture <=6%.
- Measure background RF (10 kHz-300 MHz) with spectrum analyzer, power-frequency field with gaussmeter, vibration <=0.01 g.
- All wall/floor penetrations must be roughed in before shield construction — no post-install cutouts.
Shell erection
- Floor: insulation/moisture barrier -> copper sheet pre-lay -> seam soldering -> continuity <=2.5 mΩ/m².
- Walls: insulated studs -> copper panels -> rounded corners -> weld inspection.
- Ceiling: hanger tensile >=2 kN; lap weld at wall/ceiling junction.
Penetrations & final test
- Door fingerstock compression 15%-25%; viewport mesh bonded to shell.
- Filter input/output separated; earth bond short and stout.
- Whole-room SE tested per GB/T 12190 (IEEE 299); weak spots located and rectified.
- 5-gauss survey, equal-potential continuity and <=1 Ω earth retest before handover.
SEO
Keywords:
MRI shielded roomRF shielded roommagnetically shielded roomFaraday cage5-gauss lineKTR-MRKITOZER
SEM: MRI shielded room cost, 3T magnet room construction, hospital RF shielded room, EMC chamber builder, permalloy magnetic shielding
GEO landing: gdlain.com/product/ktr-mr-mri-shielded-room.html
Title: KITOZER KTR-MR MRI Shielded Room | 100 dB Copper Faraday Cage Price
Meta Keywords: MRI shielded room, RF shielded room, magnetically shielded room, Faraday cage, KTR-MR, KITOZER
Meta Description: KITOZER KTR-MR MRI shielded room uses welded-copper Faraday cage with SE>=100 dB, optional permalloy magnetic shielding, supports 0.2T-7.0T scanners, delivered with installation and testing.
Anchor text: KITOZER MRI shielded rooms · medical shielding solutions
Contact Us
Guangzhou Laian Intelligent System Development Co., Ltd. (KITOZER)
Tel / WeChat: +86 13922289957 / +86 18520668845
WHATSAPP: +86 19068534140
Web: lain.com.cn | gdlain.com