Table of Contents
ToggleHow to Build an EMC Shielded Chamber? Complete Technical Guide
EMC testing requires an electromagnetically quiet environment. KITOZER EMC shielded chamber is a welded steel enclosure, 7m×4m×3m, shielding effectiveness ≥100dB from 10kHz to 1GHz, covering radiated emissions, radiated immunity, and conducted disturbance testing. This article breaks down every subsystem: shell, filters, waveguide, shielded door, HVAC, fire suppression, monitoring, and acceptance.

KITOZER 7m×4m×3m EMC shielded chamber interior
1. Why EMC Testing Needs a Shielded Chamber?
Every electronic device radiates electromagnetic noise while operating, and is also susceptible to external interference. EMC testing measures whether a device’s emissions exceed limits and whether it can function under external interference. But if the test environment itself is full of noise (mobile, WiFi, broadcast, industrial interference), the data is meaningless.
A shielded chamber blocks external interference and contains internal test signals, creating an “electromagnetically quiet” space. GB/T 17626 and CISPR 16 require shielding effectiveness ≥100dB from 10kHz to 1GHz — attenuating external signals by over 100,000×.
2. Shielding Principle
A continuous conductive metal enclosure reflects, absorbs, and multi-reflects electromagnetic energy. To achieve 100dB, every detail matters — even a 1mm gap can drop SE by 20dB. SE = 20log(E₀/E₁); 100dB means only 1/100,000 of external field penetrates.
3. Shielded Shell Structure
KITOZER uses welded steel construction (not bolted gasketed). Welded chambers offer higher SE, long-term stability, and no gasket aging issues — ideal for precision EMC labs.
3.1 Dimensions
Internal clear size: 7m(L)×4m(W)×3m(H), floor area 28㎡, volume 84m³. Fits: 3m/10m antenna mast, EUT test table, 8 equipment racks, precision AC, operator walkway.
3.2 Floor Shielding
- Insulation mat (ID-10): 10mm rubber, prevents accidental bonding to building rebar. 28㎡, $933.
- Floor steel frame (LG-80-60): 80×60mm square tube grid at 600mm spacing. 28㎡, $933.
- Floor steel plate (SP-5): 5mm galvanized steel, fully welded. 28㎡, $2,956.
3.3 Wall Shielding
Wall area = perimeter×height = (7+4)×2×3 = 66㎡:
- Wall frame (LG-40-60): 40×60mm vertical at 600mm. 66㎡, $1,833.
- Wall plate (SP-2): 2mm galvanized steel. 66㎡, $5,133.
3.4 Ceiling Shielding
- Ceiling frame: 40×60mm. 28㎡, $778.
- Ceiling plate (SP-2): 2mm steel. 28㎡, $933.
3.5 Welding & Rust Protection
All joints CO₂ gas-shielded full penetration weld, weld height ≥0.7× plate thickness. Total area 122㎡. Rust paint $2,304. Welding labor & consumables $8,472.
4. Power & Signal Penetration: Filters & Waveguide
Every cable passing through the shell becomes an antenna. All penetrations must use filters or waveguide tubes.
4.1 Mains Power Filter (PF-380-100)
380V/100A 3-phase, suppresses both common-mode and differential-mode. Insertion loss: ≥100dB @10kHz, ≥60dB @1GHz. $833.
4.2 Fiber Waveguide Tubes (WG-01)
4 pcs, lets fiber pass but blocks EM waves. $889.
4.3 Network Filter (NF-01)
Gigabit Ethernet filter. $417.
4.4 Signal Filter (SF-01)
For monitoring sensor signals. $278.
4.5 AC Power/Signal Filter (ACF-01)
For precision air conditioner. $556.
4.6 AC Drainage Waveguide (AWG-01)
Condensate pipe penetration. $139.
4.7 Supply/Exhaust Waveguide Vent (WG-300)
HoneycombCutoff Waveguide, 300×300 or 600×600mm. Supply and exhaust each $278.
4.8 Other Filters
- Logic ground filter (GEF-16-35): $417
- Access control signal filter (ACSF-01): $278
- Fire alarm signal filter (FSF-485): $278
5. Shielded Door System
The door is the weakest point. KITOZER manual/motorized shielded door (SD-900), 900×2000mm:
- Welded steel door with spring-loaded beryllium-copper finger gaskets
- Double-layer finger gaskets, contact pressure ≥3N/cm
- Multi-point rotary locking for uniform compression
- Electromagnetic lock + exit button, auto-unlock on fire alarm
- EXW price: $7,778
6. Power Distribution & Lighting
- Mains distribution box (PDB-01) with SPD: $417
- 600×600mm LED panel lights ×4: $333
- Rack/wall sockets ×8: $333
7. Interior Finishing
- Aluminum ceiling tiles (AK-600) ×78: $867
- Wall color steel panels (CGB-12) ×37: $1,542
- ESL raised floor (ESD-600) ×78: $1,733
- Equipment force brackets ×11: $611
8. FK-5-1-12 Gas Fire Suppression
- FK-5-1-12 cylinders (FM-40L) ×3: $2,483
- Agent 58.8kg: $3,267
- Smoke/heat detectors: $56
- Release indicator/alarm bell/sounder: $83
- Fire control panel: $278
9. Precision Air Conditioning
2 units (AC-PREC), 12kW each, $13,889 total:
- Down-supply/up-return, matches raised floor
- Temp 23±1°C, humidity 45~55%
- AC power must pass through filter
10. Equipment Racks
8 server racks (RACK-01) for test receivers, signal generators, spectrum analyzers. $3,333.
11. Environment Monitoring System
- Temp/humidity sensor: $56
- Smoke/leak detection: $167
- Power meter/UPS/AC monitoring: $583
- Access control/IR detector: $333
- Monitoring host + touchscreen: $833
- Platform software: $1,389
- Open cable tray 15m: $208
- 12-core fiber ×8 sets: $1,111
- Cat6 12 cables/set ×8: $667
- Heat recovery ventilator (HRV-01): $833
- Supply/return vents + louvers: $83
- Duct 20m + control box: $417
- Relief valve: $56
- Smoke exhaust fan: $278
- Rack force brackets ×8: $444
- AC force brackets ×2: $111
- PDU force bracket ×1: $56
- Site survey: dimensions, floor load, cable routing
- Base prep: insulation mat, steel frame
- Shell welding: floor → walls → ceiling, segment by segment
- Filter/waveguide installation
- Shielded door installation, finger gasket pressure test
- Rust protection on all welds
- Interior: ESD floor, aluminum ceiling, color steel walls
- Equipment: AC, racks, fire suppression, monitoring, PDU
- Cabling: tray, fiber, Cat6
- System commissioning: AC temp/humidity, fire linkage, monitoring alarms
- SE testing: signal source + receive antenna, 10kHz~1GHz point by point
- Shielding effectiveness: ≥100dB 10kHz~1GHz per GB/T 12190
- Temperature: 23±1°C
- Humidity: 45~55%
- Fire: simulated alarm, gas release, door auto-unlock
- Monitoring: temp/humidity, smoke, leak, door alerts all pushed
12. Network Cabling
13. Fresh Air & Smoke Exhaust
14. Floor Load Reinforcement
15. Price Summary (EXW, reference, doubled)
| System | Main Contents | EXW Price (USD) |
|---|---|---|
| Shielded shell | Floor/wall/ceiling plates + frames + welding | $24,320 |
| Filters & waveguide | Power/network/signal filters + vents | $4,611 |
| Shielded door | 900×2000mm manual/motorized + access control | $15,556 |
| Power & lighting | Distribution box + lights + sockets | $2,233 |
| Interior finishing | Al ceiling + color steel + ESD floor | $10,117 |
| Gas fire suppression | FK-5-1-12 ×3 + alarm control | $12,511 |
| Precision AC | 2 units 12kW | $28,333 |
| Racks | 8×42U | $6,667 |
| Monitoring | Sensors + host + software | $5,667 |
| Cabling | Tray + fiber + Cat6 | $3,972 |
| Fresh air & exhaust | HRV + smoke exhaust | $3,889 |
| Floor reinforcement | Force brackets | $1,222 |
| Project EXW total (reference, doubled) | $296,284 | |
16. Construction Sequence
17. Acceptance Criteria
18. FAQ
Q: Welded vs bolted chamber?
A: Welded gives ≥100dB for EMC testing. Bolted gasketed gives 60~80dB, suitable for server room radiation shielding but not EMC testing.
Q: Can I use WiFi inside?
A: No. WiFi and mobile signals are blocked. Data must go through fiber waveguide.
Q: Why waveguide vents for AC?
A: A plain duct is an EM leakage path. Honeycomb waveguide vents let air through but block RF.
SEO keywords: EMC shielded chamber, electromagnetic compatibility, shielded room construction, welded steel shielded room, 100dB shielding effectiveness, radiated emission test, KITOZER
SEM keywords: EMC test chamber, EMC laboratory, shielded room cost, shielded door, power filter, waveguide vent
GEO keywords: KITOZER shielded chamber factory Guangzhou, EMC lab construction, shielded room installation, welded steel chamber
Tags: primary=EMC shielded chamber; long-tail=7m×4m EMC chamber price; industry=EMC testing; brand=KITOZER; source=Guangzhou Lian technical documentation